Olympus OmniScan MX2 Ultrasonic Flaw Detector
This second generation OmniScan MX2 increases testing efficiencies, ensuring superior manual and advanced AUT application performance with faster setups, test cycles, and reporting, in addition to universal compatibility with all phased array and ultrasound modules: past, present and future. Designed for NDT leaders, this high-end, scalable platform delivers true next-generation NDT performance.
The OmniScan MX2 offers a high acquisition rate and new powerful software features for efficient manual and automated inspection performance"”all in a portable, modular instrument.
Powerstart your day the right way with the OmniScan MX2. The OmniScan MX2 simplifies and speeds up the setup process with its intuitive Weld Overlay software feature, so you can start testing immediately. Featuring the industry-standard phased array user interface with faster-than-ever performance, a bigger and brighter 10.4 inch screen, new and unique intuitive touch-screen capabilities, and faster data transfer, enabling you to get to your next inspection quicker.
State of the Art PA2 and UT2 Modules - As a world leader in phased array technology, Olympus has just released a new line of modules compatible with MX2 instrument.
PA2 - The new phased array offer"”led by the new innovative PA2 modules, features multiple improvements, such as:
Best Phased Array and TOFD Signal Quality Ever
Better signal-to-noise ratio
More powerful pulsers
64 pure gray tone
More Multi-group CapabilitiesBright, Large Touch Screen
Ability to use PA and UT channels simultaneously
General Hardware Improvements
Higher operating temperatures (45 °C)
New OmniScan probe connector with quick latch system
Designed for IP66 environmental rating
Extended autonomy on batteries
UT2 - The new Conventional Ultrasound module features the same technology of UT channels as the PA2 modules, but offers twice as many channels.
Features:
Bright Large Touch Screen
New Phased Array and TOFD Modules
New NDT SetupBuilder Design Software
New OmniPC Analysis Software
Storage devices | SDHC card most standard USB storage devices or fast Ethernet |
Data file size | 300 MB |
USB ports | 3 |
Speaker out | Yes |
Video output | Video out (SVGA) |
Ethernet | 10/100 Mbps |
Encoder | 2-axis encoder line (quadrature up down or clock/direction) |
Digital input | 4 digital TTL inputs 5 V |
Digital output | 4 digital TTL outputs 5 V 15 mA |
Acquisition on/off switch | Remote acquisition enabled TTL 5V |
Power output line | 5 V 500 mA power output line (short-circuit protected) |
Alarms | 3 TTL 5 V 15 mA |
Analog output | 2 analog outputs (12 bits) ±5 V in 10 kO |
Pace input | 5 V TTL pace input |
Display size | 26.4 cm (10.4 in.) (diagonal) |
Resolution | 800 pixels x 600 pixels |
Brightness | 700 cd/m2 |
Number of colors | 16 million |
Type | TFT LCD |
Battery type | Smart Li-ion battery |
Number of batteries | 1 or 2 (battery chamber accommodates two hotswappable\nbatteries) |
Battery life | Minimum 7 hours with two batteries |
Operating temperature\nrange | -10 °C to 45 °C\n(14 ºF to 113 ºF) |
Storage temperature range | 20 °C to 60 °C (4 ºF to 140 ºF) with batteries\n20 °C to 70 °C (4 ºF to 158 ºF) without batteries |
Relative humidity | Max. 70% RH at 45°C noncondensing |
Ingress Protection Rating | IP54 |
Shockproof rating | Drop-tested according to MIL-STD-810G 516.6 |
Connectors | 1 Phased Array connector: Olympus PA connector\n2 UT connectors: LEMO 00 |
Number of focal laws | 256 |
Probe recognition | Automatic probe recognition |
Aperture | 32 |
Number of elements | 128 |
PULSER | PA CHANNELS |
Voltage | 40 V 80 V and 115 V |
Pulse width | Adjustable from 30 ns to 500 ns resolution of 2.5 ns |
Pulse shape | Negative square wave |
Output impedance | < 25 O |
RECEIVER | PA CHANNELS |
Gain | 0 dB to 80 dB maximum input signal\n550 mVp-p (full-screen height) |
Input impedance | 65 O |
System bandwidth | 0.6 MHz to 18 MHz\n(3 dB) |
Scan type | Sectorial and linear |
Digitizing frequency | 400 MHz (12 bits) after interpolation per 4* |
Group quantity | Up to 8 |
Maximum pulsing rate | Up to 10 kHz (C-scan) |
Number of data points | Up to 8192 |
Real-time averaging | PA: 2 4 8 16\nUT: 2 4 8 16 32 64 |
Rectifier | RF full wave half wave + half wave |
Filtering | PA Channel: 3 low-pass 3 band-pass and 5 high-pass\nfilters.\nUT channel: 3 low-pass 6 band-pass and 3 high-pass\nfilters (8 low-pass filte |
Video filtering | Smoothing (adjusted to probe frequency range) |
A-scan refresh rate | Real time: 60 Hz |
On internal clock | 1 Hz to 10 kHz |
On encoder | On 2 axes: from 1 to 65536 steps |
Number of points | 32: One TCG curve per focal law |
Number of alarms | 3 |
Conditions | Any logical combination of gates |
Analog outputs | 2 |
Dimensions | US: 8.9 x 7.2 x 1.6 in. Metric: 226 x 183 x 40 mm |